The HVC80804 is a compact, safe, low-leakage, and electrically stable integrated X-ray source. It can operate continuously at 350W within the allowable temperature range. The system includes a high-voltage power supply, filament power supply, X-ray tube, high-voltage oil tank, and radiator. This high-frequency, self-cooling, and self-protecting X-ray generator delivers up to 80kV and 350W. It primarily consists of a control box and high-voltage oil tank, with RS232 interface for system control, monitoring, and firmware upgrades.
Food testing, Shoes and clothing detection, fluorescence analysis application and other fields, mostly used for X-ray machines.
| Item | Specification | 
| Input voltage | 230VAC±10%, 50/60Hz, 3Amps | 
| Output power of X ray tube | Max continuous output power 350W(80kV/4.3mA or 43kV/8.0mA | 
| Output voltage | Rated output voltage:Continuously adjustable voltage range 30kV—80kV | 
| Output voltage ripple:±0.5%(peak to peak) | |
| Output voltage accuracy:±2% of voltage setting value | |
| line regulation:±0.1% | |
| load regulation:±0.1% | |
| Tube current | Rated tube current:Continuously adjustable current range 0.5mA-8.0mA | 
| Tube current accuracy:±0.2mA of current setting value | |
| line regulation:±0.1% | |
| load regulation:±0.1% | |
| Rise time of output voltage | The kV rise time is <0.6 Sec at maximum power | 
| The kV rise time is <0.1 Sec at low voltage ( < 40kV ) | |
| Filament power supply: | input voltage:24VDC | 
| filament voltage:2.0 to 5.3Vac | |
| filament:3.0 to 3.8 Amps RMS | |
| preheating time:3sec | |
| Tube feature | Tube type:fixed anode、 beryllium window tube 、tungsten target | 
| focus:0.8mm | |
| inherent filtration:0.8mm Be | |
| radiation angle:45ºcone beam | |
| target angle:25° | |
| Cooling | transformer oil,external circulation,external forced air cooling | 
| Working temperatures | -10℃—40℃ | 
| Storing temperature | -20℃—60℃ | 
| System temperature protection | 60℃±3℃ of oil temperature | 
| Humidness | 98%,Non-condensation | 
| Weight | Oil tank:24.8kg | 
| Control Unit:3.8kg | |
| Installation direction | Installation in any direction | 
| Radiation angle | cone angle of ray tapered beam 45° | 
| X-ray leakage | Less than 0.5mR/hr at 5cm from the surface of the HVC80804. | 
JB1/AC~(AC Input Power Connector)
| Pin | Signal | Parameter | 
| 1 | L | live wire | 
| 2 | N | Neutral line | 
| 3 | G | PE | 
JB2/COM (DMR-9S interface definitions)
| Pin | Signal | Parameter | 
| 1.4.6.7.8.9 | N/C | No connect | 
| 2 | TXD | Data transmit | 
| 3 | RXD | Data receive | 
| 5 | GND | Signal gnd | 
JB3/Interlock,(DMR-9P interface definitions)
| Pin | Signal | Parameter | 
| 3/4/5/6/7/8/9 | N/C | No connect | 
| 1 | Interlock Out | |
| 2 | Interlock In | 
Short connect pin1 and pin2 make X ray source normal operation. Typical connection:
Led indicator:
| ID | Color | Meaning | 
| XrayOn | Yellow | indicate X ray on | 
| ARC | Red | Arcing in oil tank | 
| OT | Red | Over temperature | 
| EP_Err | Red | Tube voltage error | 
| IP_Err | Red | Tube current error | 
| Power | Green | Power on | 
Tank size

Copyright@2024, All Rights Reserved