The HVP1102001 is a compact, portable, and reliable high-voltage X-ray generator designed for head imaging equipment. It can continuously operate at up to 1200W for 25 seconds within the specified temperature range, relying on self-cooling without the need for additional cooling. This X-ray source can output 110kV/1200W in continuous mode (1700W in pulse mode) and features high frequency, self-cooling, self-protection, and independent control. It consists of an X-ray high-voltage oil tank and a control box, with control, status monitoring, and programming managed through a DB9 interface. Additionally, the HVP1102001 supports both continuous and pulse operation modes.
Dental X-ray imaging equipment (CT and panoramic), head scan and mobile imaging equipment
| Item | Specification | 
| Input voltage | 230VAC±10%, 50/60Hz, 9.2A,PF>0.98. | 
| Output power of X ray tube | Maximum continuous output power 1200W up to 25s (100kV, 12mA)
 Maximum pulse output power 1700W up to 50s (100kV, 17mA)  | 
| Output voltage | Rated output voltage:Continuously adjustable voltage range 50kV—110kV | 
| Output voltage ripple:±0.5%(peak to peak) | |
| Output voltage accuracy:±1% of voltage setting value | |
| line regulation:±0.1% | |
| load regulation:±0.1% | |
| Tube current | Rated tube current:Continuous mode: adjustable from 2 to 15mA (Max power 1200W); Pulse mode: adjustable from 2 to 20mA (maximum power 1700W). | 
| Tube current accuracy:±0.1mA of current setting value | |
| line regulation:±0.1mA | |
| load regulation:±0.1mA | |
| Filament power supply: | input voltage:24VDC | 
| filament voltage:4.0 to 8.0Vac | |
| filament:2.2 to 4.0 Amps RMS | |
| preheating time:3sec | |
| Tube feature | Tube type:fixed anode、glass envelope、tungsten target | 
| focus:0.5mm | |
| inherent filtration:1.5mm Al | |
| radiation angle: 30º, cone beam | |
| target angle:5° | |
| Operation mode | continuous mode, pulse mode | 
| Pulse frequency | 20-50Hz | 
| Pulse time | Single pulse exposure time:5.0~50ms | 
| Dual energy setting | Arbitrary setting of two different energy voltage levels of tube voltage within the rated range | 
| Max exposure time | Maximum continuous exposure time 25s at continuous mode | 
| Maximum continuous exposure time 50s at pulse mode | |
| Kv Rise Time at maximum power: | Continuous mode:To 100kV and 10mA in less than 15ms | 
| Pulse mode:rise time of single pulse kV less than 1ms | |
| Cooling | transformer oil,aluminum shell natural heat dissipation | 
| Working temperatures | -10℃—40℃ | 
| Storing temperature | -20℃—60℃ | 
| System temperature protection | 60±3℃ of oil temperature | 
| Humidness | 10%-95%,Non-condensation | 
| Weight | X-ray tank: 15.0kg | 
| Control box: 3.6kg | |
| Installation direction | Installation in any direction | 
| Radiation angle | cone angle of ray tapered beam 30° | 
| X-ray leakage | Less than 100mR/hr @1meter from the surface of the HVP1102001 Integrated X-Ray Source. | 
JB1/AC~(AC Input Power Connector)
| Pin | Signal | Parameter | 
| 1 | L | live wire | 
| 2 | N | Neutral line | 
| 3 | G | PE | 
JB2/COM(Communication Interface, DMR-9S(female)
| Pin | Signal | Parameter | 
| 1.4.6.7.8.9 | N/C | Undefined | 
| 2 | TXD | Data transmit | 
| 3 | RXD | Data receive | 
| 5 | GND | Reference ground | 
JB3 (Signal Interface, DMR-9P(male)
| Pin | Signal | Parameter | 
| 1 | Interlock Out | |
| 2 | Interlock In | |
| 3 | Syn_ Out | synchronizing signal output | 
| 4 | Syn_ In | synchronizing signal input | 
| 5 | GND | Reference gnd | 
| 6 | Syn_Out | synchronizing signal output | 
| 7 | GND | Reference ground | 
| 8 | N/C | Undefined | 
| 9 | Xray_ON | X-ray switch signal pin. | 
Short connect pin1 and pin2 make X ray source normal operation. Typical connection
Tank size