The HVD801B1 X-ray source is a compact, low-leakage, and electrically stable high-voltage X-ray generator which operates in two modes: single-energy and dual-energy. Within the permissible operating temperature range, the X-ray source can continuously operate in single-energy mode at a maximum power of 80W for 10 seconds. In dual-energy mode, it can switch continuously between two energy levels, with adjustable voltage, current, and operating time for both levels. The X-ray source consists of a control unit, high-voltage components, and an X-ray tube. This X-ray source features high frequency, self-cooling, and high integration, with a maximum voltage of 80kV and a maximum power of 80W. It is primarily composed of a control box and a high-voltage X-ray module. The X-ray source utilizes RS232 interface for system control, monitoring, and firmware upgrades.
Mobile X-ray testing equipment in the field of human bone density.
Item | Specification |
Input voltage | 22V~26V,15Amps |
Output power of X ray tube | Max continuous output power 80W(80kV/1mA) |
Output voltage | Rated output voltage:Continuously adjustable voltage range 30kV—80kV |
Output voltage ripple:±0.5%(peak to peak) | |
Output voltage accuracy:±2% of voltage setting value | |
line regulation:±0.1% | |
load regulation:±0.1% | |
Tube current | Rated tube current:Continuously adjustable current range 0.1mA-1.0mA |
Tube current accuracy:±0.01mA of current setting value | |
line regulation:±0.1% | |
load regulation:±0.1% | |
Rise time of output voltage | Continuous mode voltage rise time: under the condition that the output voltage is less than 80kV, the output voltage rise time is less than 0.1 seconds; |
Voltage rise time in dual-energy mode: Under dual-energy operating conditions, the output voltage rise time is less than 0.1 seconds. | |
Filament power supply: | input voltage:24VDC |
filament voltage:2.0 to 5.3V | |
filament:3.0 ± 0.2 Amps RMS | |
preheating time:3sec | |
Tube feature | Tube type:fixed anode、 glass envelope 、tungsten target |
focus:0.8mm | |
inherent filtration:1.8mm AI | |
radiation angle:40ºcone beam | |
target angle:20° | |
Cooling | Electronic silicone solid-state packaging, natural heat dissipation. |
Working temperatures | -10℃—40℃ |
Storing temperature | -20℃—60℃ |
System temperature protection | 60℃±3℃ of oil temperature |
Humidness | 98%,Non-condensation |
Weight | Split type X-ray source :4kg |
Installation direction | Installation in any direction |
Radiation angle | cone angle of ray tapered beam 40° |
X-ray leakage | At 1 meter of the external surface of the tank, the radiation leakage is not more than 3mR/hr. |
JB1/DC,(DC Input Power Connector)
JB2/COM, (RJ45 interface definitions)
JB3,Extern Interface(12P)
Order number | Feet function |
1 | Interlock In |
2 | Interlock Out |
3 | X-ray on relay contact common |
4 | X-ray on relay contact N/C |
5 | X-ray on relay contact N/O |
6 | / |
7 | / |
8 | / |
9 | SGND |
10 | 5V |
11 | EP |
12 | IP |
Tank size
Copyright@2024, All Rights Reserved