The HVC802B0 X-Ray Source is an exceptionally compact, portable, and reliable line of high voltage X-ray generators. Using in allowable operating ambient temperature, it is capable to self-heat dissipate up to 100W continuous with no additional cooling. The X-ray generator integrates the high voltage inverter, filament supply, X-ray tube and X-ray housing into one unit. The High Voltage X-ray Unit is an 80kV/100W, High Frequency, self-cooled, self-protected & self-contained X-ray generator. It consists of an X-ray tank, and a control box. The Generator is controlled, programmed and monitored via a standard RS232 interface.
Food testing, industrial non-destructive testing, dangerous goods testing and other fields, mostly used for simple X-ray machines or mobile X-ray testing equipment
| Item | Specification |
| Input voltage | 230VAC±10%, 50/60Hz, 1.2Amps |
| Output power of X ray tube | Max continuous output power 100W(80kV/1.25mA,50kv/2mA) |
| Output voltage | Rated output voltage:Continuously adjustable voltage range 20kV—80kV |
| Output voltage ripple:±0.5%(peak to peak) | |
| Output voltage accuracy:±2% of voltage setting value | |
| line regulation:±0.1% | |
| load regulation:±0.1% | |
| Tube current | Rated tube current:Continuously adjustable current range 0.5mA-2mA |
| Tube current accuracy:±1% mA of current setting value | |
| line regulation:±0.5% | |
| load regulation:±0.5% | |
| Rise time of output voltage | At maximum power, the output voltage rise time is less than 0.6 second |
| When the output voltage is below 40kV, the rise time is less than 0.1 seconds. | |
| Filament power supply: | input voltage:24VDC |
| filament voltage:1 to 4ac | |
| filament:1.0 to 4.5 Amps RMS | |
| preheating time:3sec | |
| Tube feature | Tube type: fixed anode, glass window, tungsten target |
| focus:0.5mm | |
| inherent filtration:0.8mm Be,0.7mm AI | |
| radiation angle:80°*16° fan beam | |
| target angle:20° | |
| Cooling | transformer oil,natural air cooling |
| Working temperatures | -10℃—40℃ |
| Storing temperature | -20℃—60℃ |
| System temperature protection | 60°C ± 3°C of Oil temperature: |
| Humidness | 98%,Non-condensation |
| Weight | 15kg |
| Installation direction | Installation in any direction |
| Radiation angle | 16°×80° |
| X-ray leakage | Less than 0.5mR/hr at 5cm from the surface of the HVC802B0. |
JB1/AC~(AC Input Power Connector)
| Pin | Signal | Parameter |
| 1 | L | live wire |
| 2 | N | Neutral line |
| 3 | G | PE |
JB2/COM (DMR-9S interface definitions)
| Pin | Signal | Parameter |
| 1.4.6.7.8.9 | N/C | No connect |
| 2 | TXD | Data transmit |
| 3 | RXD | Data receive |
| 5 | GND | Signal gnd |
JB3/Interlock,(DMR-9P interface definitions)
| Pin | Signal | Parameter |
| 3/4/5/6/7/8/9 | N/C | No connect |
| 1 | Interlock Out | |
| 2 | Interlock In |
Short connect pin1 and pin2 make X ray source normal operation. Typical connection:
Tank size