HVC802B0 X-Ray Source

Key Features:

  • Integrated design with high electrical integration and a compact appearance
  • Capable of continuous, uninterrupted operation for extended periods
  • High stability
  • Can be installed in any orientation
  • Standard digital interface, easy to use

Product Details

The HVC802B0 X-Ray Source is an exceptionally compact, portable, and reliable line of high voltage X-ray generators. Using in allowable operating ambient temperature, it is capable to self-heat dissipate up to 100W continuous with no additional cooling. The X-ray generator integrates the high voltage inverter, filament supply, X-ray tube and X-ray housing into one unit. The High Voltage X-ray Unit is an 80kV/100W, High Frequency, self-cooled, self-protected & self-contained X-ray generator. It consists of an X-ray tank, and a control box. The Generator is controlled, programmed and monitored via a standard RS232 interface.

Application:

Food testing, industrial non-destructive testing, dangerous goods testing and other fields, mostly used for simple X-ray machines or mobile X-ray testing equipment

Specification:

Item Specification
Input voltage 230VAC±10%, 50/60Hz, 1.2Amps
Output power of X ray tube Max continuous output power 100W(80kV/1.25mA,50kv/2mA)
Output voltage Rated output voltage:Continuously adjustable voltage range 20kV—80kV
Output voltage ripple:±0.5%(peak to peak)
Output voltage accuracy:±2% of voltage setting value
line regulation:±0.1%
load regulation:±0.1%
Tube current Rated tube current:Continuously adjustable current range 0.5mA-2mA
Tube current accuracy:±1% mA of current setting value
line regulation:±0.5%
load regulation:±0.5%
Rise time of output voltage At maximum power, the output voltage rise time is less than 0.6 second
When the output voltage is below 40kV, the rise time is less than 0.1 seconds.
Filament power supply: input voltage:24VDC
filament voltage:1 to 4ac
filament:1.0 to 4.5 Amps RMS
preheating time:3sec
Tube feature Tube type: fixed anode, glass window, tungsten target
focus:0.5mm
inherent filtration:0.8mm Be,0.7mm AI
radiation angle:80°*16° fan beam
target angle:20°
Cooling transformer oil,natural air cooling
Working temperatures -10℃—40℃
Storing temperature -20℃—60℃
System temperature protection 60°C ± 3°C of Oil temperature:
Humidness 98%,Non-condensation
Weight 15kg
Installation direction Installation in any direction
Radiation angle 16°×80°
X-ray leakage Less than 0.5mR/hr at 5cm from the surface of the HVC802B0.

JB1/AC~(AC Input Power Connector)

Pin Signal Parameter
1 L live wire
2 N Neutral line
3 G PE

JB2/COM (DMR-9S interface definitions)

Pin Signal Parameter
1.4.6.7.8.9 N/C No connect
2 TXD Data transmit
3 RXD Data receive
5 GND Signal gnd

JB3/Interlock,(DMR-9P interface definitions)

Pin Signal Parameter
3/4/5/6/7/8/9 N/C No connect
1 Interlock Out
2 Interlock In

Short connect pin1 and pin2 make X ray source normal operation. Typical connection:

 