The HVD801B1 X-ray source is a compact, low-leakage, and electrically stable high-voltage X-ray generator which operates in two modes: single-energy and dual-energy. Within the permissible operating temperature range, the X-ray source can continuously operate in single-energy mode at a maximum power of 80W for 10 seconds. In dual-energy mode, it can switch continuously between two energy levels, with adjustable voltage, current, and operating time for both levels. The X-ray source consists of a control unit, high-voltage components, and an X-ray tube. This X-ray source features high frequency, self-cooling, and high integration, with a maximum voltage of 80kV and a maximum power of 80W. It is primarily composed of a control box and a high-voltage X-ray module. The X-ray source utilizes RS232 interface for system control, monitoring, and firmware upgrades.
Mobile X-ray testing equipment in the field of human bone density.
| Item | Specification | 
| Input voltage | 22V~26V,15Amps | 
| Output power of X ray tube | Max continuous output power 80W(80kV/1mA) | 
| Output voltage | Rated output voltage:Continuously adjustable voltage range 30kV—80kV | 
| Output voltage ripple:±0.5%(peak to peak) | |
| Output voltage accuracy:±2% of voltage setting value | |
| line regulation:±0.1% | |
| load regulation:±0.1% | |
| Tube current | Rated tube current:Continuously adjustable current range 0.1mA-1.0mA | 
| Tube current accuracy:±0.01mA of current setting value | |
| line regulation:±0.1% | |
| load regulation:±0.1% | |
| Rise time of output voltage | Continuous mode voltage rise time: under the condition that the output voltage is less than 80kV, the output voltage rise time is less than 0.1 seconds; | 
| Voltage rise time in dual-energy mode: Under dual-energy operating conditions, the output voltage rise time is less than 0.1 seconds. | |
| Filament power supply: | input voltage:24VDC | 
| filament voltage:2.0 to 5.3V | |
| filament:3.0 ± 0.2 Amps RMS | |
| preheating time:3sec | |
| Tube feature | Tube type:fixed anode、 glass envelope 、tungsten target | 
| focus:0.8mm | |
| inherent filtration:1.8mm AI | |
| radiation angle:40ºcone beam | |
| target angle:20° | |
| Cooling | Electronic silicone solid-state packaging, natural heat dissipation. | 
| Working temperatures | -10℃—40℃ | 
| Storing temperature | -20℃—60℃ | 
| System temperature protection | 60℃±3℃ of oil temperature | 
| Humidness | 98%,Non-condensation | 
| Weight | Split type X-ray source :4kg | 
| Installation direction | Installation in any direction | 
| Radiation angle | cone angle of ray tapered beam 40° | 
| X-ray leakage | At 1 meter of the external surface of the tank, the radiation leakage is not more than 3mR/hr. | 
JB1/DC,(DC Input Power Connector)
JB2/COM, (RJ45 interface definitions)
JB3,Extern Interface(12P)
| Order number | Feet function | 
| 1 | Interlock In | 
| 2 | Interlock Out | 
| 3 | X-ray on relay contact common | 
| 4 | X-ray on relay contact N/C | 
| 5 | X-ray on relay contact N/O | 
| 6 | / | 
| 7 | / | 
| 8 | / | 
| 9 | SGND | 
| 10 | 5V | 
| 11 | EP | 
| 12 | IP | 
Tank size